[Product Name]
Dilution solution for blood cell analysis
[Packaging Specifications]
ZXCELLPACK-DFL 1.5L×2;1.5L×1;1L×1
[Expected use]
It is used to dilute whole blood samples to be tested, and cooperate with hematology analyzers or reagents for the detection of analytes. Clinically, it is used to prepare cell suspensions by sample dilution before blood cell analysis.
[Principle of Inspection]
Dilution solution is used to spheroidize red blood cells during the process of blood cell detection, promoting rapid staining of reticulocytes using staining components in the staining solution for blood cell analysis. Combined with laser scattering and fluorescence staining flow cytometry technology, reticulocytes are detected.
[Storage conditions and expiration date]
1. Storage conditions: 2-35 ℃, relative humidity not exceeding 90%, no corrosive gases, and well ventilated environment.
2. Validity period: unopened reagent, stored normally, with a validity period of 18 months; The validity period after opening is 60 days.
[Applicable instruments]
SYSMEX XN series blood cell analyzer
[Product Performance Indicators]
1. Appearance: This product is a clear liquid without sediment, particles, or flocs
2. pH value: 9.10 ≤pH≤ 9.50
3. Conductivity: 11.70≤p≤12.70 (ms/cm)
4. Osmotic pressure: 208≤n≤228 (mOsm/kg)
[Precautions]
1.Please confirm if there is any contamination or color change before use. Once discovered, do not use.
2.During the validity period, the diluent must be placed in a clean environment at a suitable temperature of 15-30℃.
3.If there is any freezing phenomenon, please thaw it and mix it evenly before use.
4.Please take personal protective measures during the operation.
5.This product is only for external diagnostic use.